![]() |
Frequently asked Questions |
Downloading, Viewing and Saving Data from a LogTag
LogTag Analyser Error Messages
Q: I am having difficulties communicating to my
USB interface.
A: Driver installation with USB devices can sometimes
have problems that are outside the control of LogTag Recorders. We would advise
to read through the special
manual on USB installation available for download from the LogTag website.
Occasionally the drivers de-register themselves,
which is an issue of the operating system. In this case, simply unplug the USB
interface and plug it back again (for at least 10 seconds, this should resolve
the issue.
It is highly unlikely that a USB interface
is faulty.
Q: The Installer reports the drivers are not certified
by Microsoft – what do I do?
A: LogTag recorders is
in the process of having the USB drivers signed by Microsoft. Until then simply
continue the installation by clicking the “install anyway” button.
Q: The LogTag Analyser Installer asks for all USB
devices to be disconnected – what do I do?
A: When the LogTag Analyser Installer determines
that the USB drivers need to be updated while one or more USB interface cradles
are connected, it will prompt the user to disconnect them.
You will only need to remove your LogTag USB equipment, having other USB devices
like Mouse, keyboard or memory sticks connected during this process is OK. Once
you have disconnected the USB interface cradles, the "Next >" button
in the installer will allow you to continue with the installation.
However, if the USB drivers do not
successfully unload, because for example the Analyzer software is running, when
you click the "Next >" button, the installation software will prompt
you to disconnect the USB interface cradles again. To force the USB drivers
to unload, please follow the manual removal steps described here.
Q: How do I uninstall my LogTag
Interface cradle USB Interface drivers completely?
A1: To uninstall USB interface drivers completely,
following steps are required for Windows 2000, Windows XP and Windows Vista:
1.
With USB interface cradle plugged into the computer,
Open “Control Panel” - “System”
Select the “Hardware” tab in “System Properties” and select the “Device Manager”
button
Scroll to the “Universal Serial Bus controllers” entry, expand the list by clicking
on the “+” symbol”. Right click any “USB Interface” entries and select the “Uninstall”
command from the popup context menu.
2.
Open “Control Panel” – “Add/Remove Programs”, highlight
the LogTag USB interface drivers and select “remove”
3.
Delete usbint.inf and usbint.pnf from the Windows/Inf folder
4.
Delete ftdibus.sys from Windows/system32/drivers folder
5.
Delete all DLL’s starting with “ftdi” from Windows/system32
folder.
A2: To uninstall USB drivers completely, following
steps are required for Windows 98, Windows 98SE and Windows Me:
1.
With USB interface cradle plugged into the computer,
Open “Control Panel” – “System”,
Select the “Device Manager” tab in the “System Properties”
2.
Open “Control Panel” – “Add/Remove Programs”, highlight
the interface drivers and select “remove”
3.
Delete the files “usb_ic.inf” from the “Windows/inf/other”
folder
4.
Delete ftdibus.sys from Windows/system32/drivers folder
Software Installation/Uninstallation
issues
Q: How can I uninstall the LogTag Analyser completely?
A: Open “Control Panel” – “Add/Remove Programs”,
highlight the LogTag Analyser section and select “remove”
On some occasions this may not remove every single file from your computer.
In that case please delete following files also:
- all LogTagIO14*.dll files in #Windows#\System32 folder
- userprofile.dat file from C:\Documents and Settings\#USER#\Application Data\LogTag
Q: HELP – I have accidentally switched to a different
language and now I don’t know where I am!
A: Exit the LogTag Analyser software and delete
the file “userprofile.dat” from C:\Documents and Settings\#USER#\Application
Data\LogTag. You may need to change the display settings in Windows Explorer
to be able to navigate to this file. Please refer to the Windows Operating System’s
online help. Once this file is deleted, the software menus will be displayed
in English or the same language of the operating system, if LogTag Analyser
includes support for that language. Please note, that other settings you may
have made like Communication port settings or default chart parameters are also
lost.
Q: My Wizard reports “COM1, COM2, COM3, COM4 not
found”, what do I do?
A: This error means that LogTag Analyser can not
find an interface plugged into any of these ports. If you are using an RS232
interface, and it is not found by the software, the most likely cause is the
plug not being inserted correctly.
If you use a USB interface, and you do not wish to see this error message all
the time, simply go to “Options” – “Communication Ports” and remove the tick
in front of the Communication ports you are not using. This will also speed
up your download process as the software no longer uses these ports.
Q: I am getting an error message “Internal error:
$4:USB.0012345” – what does it mean?
A: “Internal error $4” is a very general error
message which simply means that LogTag Analyzer found an interface and a LogTag
unit inside the interface, but was unable to establish communication. In other
wors, a $4 error means there was no reply from the LogTag.
The letters and numbers after $4 represent type and serial number of the interface.
Often this error will disappear by simply repeating the action you were performing
when the error occurred, but sometimes it can be a bit more difficult to locate
the cause. Here are some scenarios in which a $4 error could occur and what
to do to rectify the issue:
-
There is poor contact between the LogTag and the interface
contacts; this can happen if the contacts are dirty or have suffered from exposure
to a corrosive environment. Usually this can be rectified by cleaning the contacts
on the LogTag (a good way is to use a soft pencil eraser.)
-
There is a “stuck” contact pin in the interface; look
inside the slot – there should be three contact pins visible, they should slide
in an out with moderate resistance when force is applied (e.g. with the end
of a pencil)
-
There is not enough power to the interface (USB); We have seen this error when USB interfaces are used on unpowered
USB hubs with other equipment also connected. USB ports have a limited ability
to supply power and we would always recommend to connect the USB interfaces
directly to the computer or to use a powered hub.
-
Communication has been interrupted prematurely; sometimes
a user will disconnect the LogTag too early, which will cause this error to
be displayed. In this case simply re-do the action with the LogTag connected
properly throughout the communication.
-
The USB port is defective; Try a different port and see
if the error persists.
-
The LogTag battery is critically low; This could happen after a long trip, and would be amplified
by trying communication if the unit came out of a cold environment. Lithium
cells and in fact all batteries do have the habit of exhibiting a lower battery
voltage in cold conditions, so downloading units that have come e.g. directly
out of a freezer could experience communication problems. In that case it may
pay to wait until the unit has acclimatised to room temperature or slightly
above.
-
There was water in the unit; individual LogTag models
have different ingress protection and are not waterproof. If there is water
ingress the unit will keep working for a time, but eventually the communication
will cease to work.
Please also
remember you should always keep your software up to date.
Q: I am getting an error message with a strange
code – what does it mean?
A: Please find a summary of error codes and solutions
here.
Q: My LogTag can not be configured!
A: Some LogTags are purchased as single trip loggers
and can not be reconfigured once used to record a trip. This will be shown by
a lock
symbol next to the USB interface in the program screen. The LogTag will need
to be returned to your distributor to have the lock removed.
Q: I can start a LogTag by pressing the start button,
how do I stop it logging?
A: Logging cannot be manually stopped on the LogTag
itself. If you want a LogTag to stop recording at a specific time, select a
specific number of readings or select a specific time period to be covered in
the “configure for next use” screen. The LogTag will then stop recording data
when that point is reached. If you choose “Record Readings Continuously”, the
oldest recording will be overwritten once the memory of the LogTag is full,
and the LogTag can only be stopped by re-configuring or hibernating
it with the LogTag Analyser Software.
Note that if you wish to indicate
when a load has safely reached its destination, you can use the “Inspection
Mark” feature instead of stopping the LogTag. When the "Start-Mark"
button is pressed while logging is in progress, an inspection mark is inserted
in the recorded data. This enables confirmation that the load has been manually
checked at certain times in the journey, and also enables the recipient of the
shipment to show when the load arrived at its destination.
Q: What happens if I press the “Again Button” with
a different type of LogTag?
A: You can reconfigure different types of LogTag
with the “Again” button, however recommend caution
when doing this. LogTag Analyser Software will only set those parameters that
are present in both LogTags. If you enter a HAXO after you just configured a
TRIX only the temperature parameters will be correctly set.
Q: What exactly is ''Pre-Start” fail-safe logging
?
A:
This function allows the LogTag to record temperature data even if the
"Start" button is not pressed.
Under normal circumstances the logger is configured using Analyser software,
then the user presses the "Start" button to initiate
data collection. This is usually done when the LogTag is placed in the consignment
of goods. However if the user forgets to press the "Start" button,
the logger will still collect data in "pre-start" mode, and this data
can be viewed in the normal way using Analyser.
Please note that Pre-Start logging can be enabled or disabled when configuring
the logger. Obviously this function has
to be enabled for Pre-Start logging to work.
Pre-start logs do not count towards the total number of readings configured.
In other words, you can not expand the memory size by utilising pre-start logging.
If a LogTag is configured to take readings up to its memory capacity and has
pre-start enabled, the pre-start readings will no longer be available once the
Start button has been pushed AND the LogTag has completed its intended full
logging cycle. The pre-start readings will be overwritten to ensure the “main”
recordings can use the complete memory.
Q: I want to start recording at a specific date
and time – but the Software won’t let me do that!
A: We are currently in final test of LogTag Analyser
1.5, which will allow this very feature.
Q: Can we use a LogTag to measure temperature of
liquids?
A: LogTags Trix-8 and HAXO-8 are not rated for
direct immersion into any liquid without the use of a waterproof bag or casing.
We do however have another version of the product, TREX-8, which is fitted with
an external sensor. This external sensor is suitable for immersion. Specifications
for this product are available on this web site
Q: Can we fit an external temperature sensor to
a TRIX-8?
A: Sorry, external temperature sensors can only
be fitted to the TREX-8.
Q: How fast will TRIX-8 react to a change in temperature?
A:
The most common industry
standard expression for temperature reading stabilisation time is the 'T90'.
This is defined as the typical time taken of a given sensor in a given environment
(i.e moving air, still air, liquid etc) to register
90% of an immediate step change in temperature.
The T90 of TRIX-8 is typically less than 5 minutes(T90)
in moving air of 1m/s.
Because thermal step response is basically dependant on the latent heat of the
two thermal masses involved (i.e the LogTag and the
surrounding environment) and the rate of thermal transfer between them, the
T90 is the same irrespective of the difference of temperatures though at extremes
(water freezing and boiling points for example) the environment does not behave
homogeneously thereby changing the rates of thermal conduction and therefore
changing the resulting T90 value.
For example: in a situation of a 25°C step change say starting at 10°C and moving
to 35°C in moving air of 1m/s, the TRIX-8 will typically register 90% of the
step after less than 5 minutes - i.e. 0.9x25 = 22.5 -> 10+22.5 = registering
32.5°C within 5mins.
The T90 for a 15°C step in 1 m/s moving air (say
10°C to 25°C) will be the same. 0.9x15 = 13.5°C +10 = registering 23.5°C
within 5mins.
Of course the T90 is different in slower moving air or still air. (though air is never totally 'still' when a temperature difference
exists as convection will take place).
Q: What is the
hibernate feature ?
A: LogTags can be switched to hibernate
mode manually in the LogTag menu.
This reduces the current consumption
on LogTags that support full hibernate to very low levels equivalent to around
1/10th or less current than normal operation. This basically extends the battery
life to around the shelf life of the battery (5-10 years).All Humidity LogTags
(HAXO-8) support this feature. LogTag Temperature (TRIX-8 & TREX-8) will
support this feature from ex-factory units shipped June 2006 onwards. Units
manufactured before this date are just put in the lowest possible power state
(no logging, no indicators blinking).
Recorders are automatically ‘woken
up’ from hibernate if they are placed in an interface and accessed and do not
automatically return to the hibernate state.
Note that recorders may wake up from
hibernate if static discharges into the contacts of the recorder as a result
of handling the recorder after it has been hibernated.
Q: How is battery condition shown and how can this be represented
as %remaining ?
A: The LogTag design incorporates a 'battery status' circuit
which measures the battery voltage and returns a 'OK'
or 'LOW' to Analyzer. This is not a physical value, just a status so it is not
possible to calculate a %remaining from this data.
However, the point at which the battery starts indicating battery LOW is typically
when the battery has around 15%-20% capacity left and therefore still has around
3-4 months of typical use remaining before totally failure.
Downloading
, Viewing and Saving Data from a LogTag
Q: I made annotations to my chart and changed the
graph title before I saved it, but now they have disappeared
– what did I do wrong?
LogTag Analyser version 1.4 can not save
any annotations directly to a file. This is a result of FDA Chapter 21 CRF Part
11 which prohibits a change to the LTD file data once downloaded from the LogTag.
We are however working on a new release of the software, version 1.5, where
the annotations will be stored in an accompanying file which will therefore
satisfy the FDA requirements and allow them to be retrieved for viewing later.
You can then also save graph titles.
Q: What causes some of the data points on the charts
to be non-validated? Or what is non-validated data?
A:
LogTag uses a very sophisticated method to automatically insert error
detection codes during recordings. This assists in detecting corruption and/or
manipulation of data. These error detection codes are saved with the file so
any attempts to manipulate data in a saved file are also detected.
Non-validated readings occur when a block of readings
fails a data integrity check; these readings are displayed differently on the
chart and data list. The most common cause for non-validated readings is when
the LogTag has been configured to do continuous recording (i.e. the new recordings
'wrap around' over previous recorded values).
As new data is over-written on a block of protected recordings, the data integrity
error check for the original block of recordings will indicate an error, thereby
technically putting the block of data into doubt.
Non-validated
data may also be displayed if there have been communications issues during data
transfer from the LogTag, or if problems occurred during logging where either
the readings or the error detection code was not correctly written to memory.
This situation may occur if the LogTag was operated in an environment far beyond
the production specification (i.e. very cold or very hot or in a heavy industrial
environment with high electro-magnetic fields present).
In very rare cases this could be a result of a memory fault in the LogTag.
Q: How is the downloaded data stored and how can this data be accessed if I send it to another person ?
A: Downloaded data is stored in LogTag LTD files
which is a proprietary secure format that is tamperproof. This is a requirement
in many applications and standards (such as the FDA 21 CFR Part 11 in the
LogTag
Analyzer has a configured folder (set in Options)
for automatic storage of downloaded data.
LTD files can
only be accessed and viewed by LogTag application
software so if another person wishes to view the data on different computer
then either they need to install LogTag Analyzer (which
is free for download) for the data is exported to another format.
LogTag
Analyzer can export data either manually or automatically to formats capable
of being imported by other applications such as Excel however exported data
is no longer secure/tamper proof.
Q: I want to view data in Microsoft Excel – how
can I export data?
A: LogTag Analyser has the option to save .CSV
files, which can be read directly into Microsoft Excel. Please see your MS Excel
documentation if you have difficulties with this step.
To create a .CSV file there are three options you can
choose from:
1. In LogTag Analyzer, go to "EDIT" - "OPTIONS"
and select "EXPORTS AND REPORTS". Tick the selection box next to "CSV
(Comma delimited)" files. From now on anytime you download a LogTag a CSV
file will automatically be saved together with the LTD file in the directory
chosen in "File and Folder settings". This is usually the "My
Documents\My LogTag Data" folder.
2. If you wish to export a previously saved file to Excel,
open that file, select "FILE" - "SAVE AS", and in the drop
down box "Save as type" at the bottom select "CSV (Comma delimited)".
The file extension will be automatically be added.
3. When you are viewing any data or
summary tab simply select “Edit” – “Copy” and paste it directly into an Excel
spreadsheet
Q: Can LogTag recorders be calibrated?
A:
There is a common misconception about the meaning of certain terminology
used in connection with calibration.
Calibration is the process
of checking a given instrument against a reference with a calibration table
of values and errors being generated. What clients usually want is the ability
to “adjust” a LogTag. Adjustment (or
recalibration as it is sometimes called) is when a
given instrument is adjusted to read in accordance with a given reference.
HAXO LogTags
can already be adjusted for RH with an application called “LogTag Calibrate”.
Please discuss this requirement with your local distributor. This software does
not currently support temperature adjustment of TRIX or TREX LogTags, but we
are working on an upgrade at the moment and hope to have this available soon.
Q: How often should LogTags be
calibrated?
A: The LogTag is designed to operate to the published
accuracy over the typical 2 year life provided it is operated in the rated environment
defined in the product specification. Temperature calibration therefore should
not be required, unless it is a third party requirement for the application
in which they are used. Due to the nature of the RH sensor it is advised that
RH calibration is performed every 6-12 months, depending on the environment
they are subject to.
Q: My LogTag is reporting an empty battery, can
I change it myself?
A: It is possible
to change the LogTag battery, but the process requires some dexterity, soldering
equipment and some electronics soldering experience.
PLEASE NOTE THAT ALL NECESSARY
SAFETY PRECAUTIONS MUST BE
The process is as follows:
1.
Hibernate the
LogTag with the LogTag Analyser software
2.
Remove front
label, remove the countersunk screws and open the case
3.
De-solder the
battery from PCB
4.
Solder new
battery in place
5.
Replace foam
dot on battery
6.
Re-assemble
case
7.
Fit new front
label
8.
Test function
You will need a replacement battery
and a replacement front label. The battery and label are custom made and must
be purchased from LogTag Recorders Ltd or an authorised distributor.
The cost of the replacement components plus the cost of labour to do this job
is typically much higher than the cost of a replacement logger. However, if you wish to proceed, you can enquire
about the cost of a replacement battery and label with your local distributor.
|
Value |
Definition |
Possible Resolution |
|
$0 |
Operation
completed successfully |
|
|
$1 |
General failure |
Obtain and
install a newer version of the software. |
|
$2 |
Unable to
communicate – No access to the communication port |
Unplug the
interface for at least 10 seconds and then re-plug interface. |
|
$3 |
File/disk
error while accessing LogTag information within a file. |
Use disk validation
and repair tools, like scandisk or chkdsk, to attempt to repair the issue. |
|
$4 |
Unable to
communicate - No Response from LogTag |
Ensure the
LogTag is fully inserted into the cradle. |
|
$5 |
Unable to
communicate – |
Check that
an interface is connected, if not disable the usage of the port in the
Options. |
|
$6 |
Memory allocation
error |
Software has
run out of RAM to operate successfully. Close some applications that are
currently running. Increase the physical amount of RAM in computer. |
|
$7 |
Invalid parameter
in communication |
An internal
error. Obtain and install a newer version of the software. |
|
$8 |
The hardware
and/or operating system of the system accessing the library method does
not support the system requirements of the library. |
Check that
the system meets minimum requirements of the software. Update IE from
Microsoft. |
|
$9 |
Password required
but not supplied |
Software was
unable to gain secure access to the LogTag. Obtain and install a newer
version of the software. |
|
$A |
Data format
error – please upgrade software |
upgrade software |
|
$B |
Unsupported
product - please update software |
upgrade software |
|
$C |
Unsupported
file – please update software |
upgrade software |
|
$D |
LogTag version
error – please update software |
upgrade software |
|
$E |
Communications
protocol error – please upgrade software |
upgrade software |
|
$F |
LogTag configuration
error- not supported or incorrect |
Obtain and
install a newer version of the software. |
|
$10 |
The software
failed to perform the operation because the user cancelled the operation
before it was completed. |
Try again. |
|
$11 |
An internal
error has occurred. |
Close all
copies of the software and/or restart computer. |
|
$12 |
LogTag Trip
Usage limit exceeded |
Return to
distributor |
|
$13 |
Unable to
communicate – |
Plug the interface
cradle into a different communication port on the computer. |
|
$14 |
User Server
cannot establish TCP connection |
Check TCP/IP
network support is installed and enabled. |
|
$15 |
User Server
cannot connect to locked user account |
Contact your
administrator to grant access to your account. |
|
$16 |
Internal error
code indicating that a component of the software is out of date. |
Obtain and
install the latest version of the software. If problem persists please
contact your distributor. |
|
$17 |
The software
is unable to successfully connect to the LogTag User Server software. |
Obtain and
install the latest version of the software. If problem persists, contact
your administrator to update the copy of LogTag User Server. |
[FAQ v8 Last updated
Last updated February 28, 2008 , © Copyright LogTag Recorders Limited