TRIX-8 Temperature Recorder Specifications

Model
  TRIX-8
Measurement Range
 
-40°C ~ +85°C
Resolution
 
< 0.1°C for -40°C ~ +40°C,
< 0.2°C for +40°C ~+80°C
Accuracy
 

better than ±0.5°C for -10°C~ +40°C.
better than ±0.7°C for -10°C~ -30°C & +40°C~+60°C
better than ±0.8°C for -30°C~ -40°C & +60°C~+80°C

Sensor reaction time
 
typically less than 5 minutes(T90) in moving air.
Capacity
 
8000 readings (16000 bytes memory)
Sampling frequency
 
adjustable, 30 sec to several hours
Recording indication
Flashing 'OK' indicator flashing 'Alert' indicator.
Download Time
Typically with full memory (8000 readings) in less than 5 secs depending on computer or readout device used.
Environmental
 
IP65
Power source
 
3V Lithium battery
Battery life
 
2 ~ 3 years of normal use (based on 15 minute logging, download data monthly.
Size
 
86mm(H)x54.5mm(W)x8.6mm(T)
Weight
 
35g
Case Material
  Polycarbonate
EMC Compliance
 

Tested and complies with EC EMC directives (EN 50081-1:1992 & EN 61000-6-1:2001)
see LogTag CE mark EMC declaration of conformance

Tested and complies FCC Part 15 Subparts A and B.
see LogTag FCC EMC declaration of conformance

EN12380 Compliance
  Tested and complies to EN12830:1999 for transport, storage and distribution of chilled, frozen, deep frozen, quick-frozen food and others.
see LogTag TRIX-8 EN12830 TÜV declaration
(Full TÜV test report number 2850509_001 available on request.)
Other features
Logging start by push button.
    Logging inspection mark in log by push button.
    Optional clearing of alert indication by push button (places inspection make at same time).
    Download inspection mark recorded in LogTag memory every time LogTag is downloaded.
    'Prestart' fail-safe logging (records temperature data even if not started)
    Low Battery indication in software.
    Comprehensive Alert indicator configuration.
    Calibration to achieve higher accuracy possible.
 
Designed for FDA CFR21 Part 11 Digital signatures
 

 

Note: Resolution & Accuracy are rated values - actual product performance may be substantially better than figures stated.

 


LogTag Recorders reserves the right to change specifications without notice.
March 2006, © Copyright Logtag Recorders Ltd 2006


Site questions or comments to webmaster@logtagrecorders.com